12VDC supply range is 9 to 14 volts.
USB supply range is 4.5 to 5.5 volts.
k) Display Device
The display device will display the complete linear operating range.
l) Environmental Stabilisation Time
The typical time to stabilise after sudden changes in environment conditions.
5 minutes after a 10
o
C change in temperature.
5 minutes after a 30% change in humidity (non-condensing).
15 seconds after a 5kPa change in ambient pressure.
m) Electric Field Strength Greater Than 10 V/m
The CEL-62X was not tested for field strengths greater than 10 V/m.
n) EMC Emissions
Emissions were no greater in any one plane or any mode of operation.
o) EMC Susceptibility
The CEL-62X has slightly higher susceptibility with the Y plane facing the
radiating antenna.
X is in the direction of the microphone, Z is in the direction of the display, and Y is
in the direction side-on to the case.